Binder CB 260 CO₂ incubator with hot air sterilization, series CB – 220-liter capacity

User Manual - Page 151

For CB 260.

PDF File Manual, 197 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 151/197
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the
charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER
service.
We recommend using the neutral cleaning agent Art. No. 1002-0016 for thorough cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER
GmbH.
NOTICE
Danger of corrosion by using unsuitable cleaners.
Damage to the chamber.
Do NOT use acidic or chlorine cleaning detergents.
Do NOT use a neutral cleaning agent on other kind of surfaces e.g., the zinc coated
hinge parts or the rear chamber wall.
To maintain the proper function of the CO
2
sensor, never spray the sensor with any cleaning detergents or
disinfectants. Generally perform any cleaning only when the chamber is turned off.
NOTICE
Damage to the CO
2
sensor due to unsuitable cleaning procedures.
NEVER spray any cleaning detergents directly on the CO
2
sensor.
Wipe the sensor with a lint-free cloth soaked with cleaning agent.
Use only the detergents or disinfectants recommended by BINDER. Other products are not approved, since
they could in particular damage the CO
2
sensor.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove any cleaning agents from the surfaces by using a
moistened towel. Let the chamber dry.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
With every decontamination method, always use adequate personal safety controls.
Following cleaning, leave the chamber door open or remove the access port plugs.
The neutral cleaning agent may cause health problems in contact with skin and if ingested.
Follow the operating instructions and safety hints labeled on the bottle of the neutral
cleaning agent.
Recommended precautions: To protect the eyes use sealed protective goggles. Wear gloves. Suitable
protective gloves in full contact with media: butyl or nitrile rubber, penetration time >480 minutes.