NORMA 6003/NORMA 6003+/NORMA 6004/NORMA 6004+
Users Manual
60
Dual-Unit Interconnection Measurement
Dual-Unit Interconnection Measurement require specific RS-485 communications parameters. These
are:
Baud Rate = 921600,
•Parity = None,
EOL = CRLF,
Data Bit = 8,
Stop bits = 1.
After the 2 Analyzers are connected through the RS-485 port, follow these steps to make
measurements:
1. Set RS-485 communication parameters of the slave.
a. From the main screen in any operating mode of the slave, select r > 2 (Comm.), and
then select the RS-485 connection method. Follow the instructions in
Set Communication
Parameters
.
b. Push J repeatedly to go back to the main screen of the Meter.
2. Set RS-485 communication parameters of the host. On the host:
a. Push r> 2 (Comm.) and then select the RS-485 connection method. Follow the
instructions in
Set Communication Parameters.
b. Push r> 3 (Remote) to enter the Remote menu.
The host’s screen prompts that the remote device is being searched, and the host will
automatically scan the connected device.
After the Analyzer as a slave is searched by the host the connected device shows in the
Available devices list, such as Address 2.
c. Push 1 (Connect) to enter the Remote menu. If the connection succeeds, the label for 2
changes to Disconnected.
d. Push J repeatedly to go back to the main screen of the System.
e. Follow the instructions in section
User Item View to enter the interface of Add Measurement
Items. Notice that 5 (Remote) is added to the category of Add measurement items.
f. Push 5 (Remote).
All user defined measurement items selected on the slave will be listed on the host’s screen.
g. Use the arrow (K L M N) keys to move the cursor to the corresponding measurement item and
push S.
The corresponding measurement items show in the list of user defined measurement items on
the host (the measurement item name has the suffix "*").
3. To exit Dual-Unit interconnection mode:
On the host, push r > 3 (Remote) > 2 (
Disconnect).
Or
On the slave, push 1 (Local).