NORMA 6003/NORMA 6003+/NORMA 6004/NORMA 6004+
Users Manual
26
Make a Wiring Selection:
1.
In the Wiring screen (
Figure 12), use the left and right arrow (M N) keys to select a channel or wiring
group in Channel Configuration. The wiring selection of the selected channel/wiring group is
highlighted.
2. Use the up and down arrow (K L) keys to select the appropriate wiring mode in Wiring Selection.
3. Push S.
4. Use the arrow keys (M N K L) and S to complete the configuration of the remaining items.
5. Push J to go back to the previous menu. Push the button continuously to go back to the main
screen.
Set Sync Source and Update Rate
The sync source is a reference input signal, and the time interval of the measurement is determined
according to the frequency of the selected sync source signal, and the sample values in this interval
are used to generate measurements such as voltage, current and power, thereby ensuring calculation
with the signal of the entire cycle to avoid measurement errors due to signal truncation.
1. From the Main screen of any view mode select
1 (Configure) > 3 (Timing Sync).
The setup interface of Timing/Sync is as shown in
Table 10.
Note
The wiring groups that can be configured in the interface depends on the wiring selection. See
Set the Wiring Selection for details.
2. Use the up and down arrow (K L) keys to move the cursor to Update Rate.
3. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
4. Use the up and down arrow (K L) keys to move the cursor to the first wiring group in Source.
5. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
Table 10. Timing and Sync
Item Description
Update rate
The time interval at which the DSP calculates the data values.
100 ms
200 ms
500 ms
•1 s
Source
The sync source of each wiring group.
All channels in each wiring group can only share the same sync source.
The available sync sources include voltage and current signals for all
channels in the wiring group.