NORMA 6003/NORMA 6003+/NORMA 6004/NORMA 6004+
Users Manual
44
Table 19. Scope Sceen
Item Description
Difference
between cursors /
average at cursor
• The difference between the 2 cursors is displayed when the double
horizontal cursors are activated.
• The average value of the cursor position is displayed when the single
vertical line cursor is activated.
• The difference between the 2 cross points of the 2 cursors and
waveform is displayed when the double vertical cursors are activated.
• If the cursor is off, no values are displayed in this area.
See
Measure with The Cursor for details.
Value at cursor
/time difference
• The values of the 2 cursors show separately when the double
horizontal cursors are activated. HIGH: Value of the upper cursor.
LOW: Value of the lower cursor.
• The maximum and minimum values of the cursors show when the
single vertical cursor is activated. MAX: Maximum value. MIN:
Minimum value.
• The time difference (t) between 2 cursors and the frequency value
(1/t) calculated accordingly are shown when the double vertical
cursors are activated.
• If the cursor is off, no values are shown in this area.
2
31
4
5
6
7
10
8
9
11