Rigol MSO7024 Oscilloscope

User Manual - Page 63

For MSO7024.

PDF File Manual, 358 pages, Read Online | Download pdf file

MSO7024 photo
Loading ...
Loading ...
Loading ...
Chapter 2 To Set the Vertical System RIGOL
MSO7000/DS7000 User's Guide 2-7
Table 2-2 Probe Ratio
Menu
Attenuation Ratio
(display amplitude of the signal: actual amplitude
of the signal)
0.01X
0.02X
0.05X
0.1X
0.2X
0.5X
1X (default)
2X
5X
10X
20X
50X
100X
200X
500X
1000X
50000X
0.01:1
0.02:1
0.05:1
0.1:1
0.2:1
0.5:1
1:1
2:1
5:1
10:1
20:1
50:1
100:1
200:1
500:1
1000:1
50000:1
Note: After the oscilloscope auto-recognized certain probes with a fixed attenuation
ratio, the probe ratio will also be auto recognized. You do not have to set it
manually.
Input Impedance
To reduce the circuit load between the oscilloscope and the circuit under test, this
oscilloscope provides two input impedance modes: 1 MΩ (default) and 50 Ω.
1 : The input impedance of the oscilloscope is very high, and the current
flowed from the circuit under test can be ignored.
50 Ω: makes the oscilloscope match with the device whose input impedance is
50 Ω.
Press 1 to open the setting menu of CH1. Then, press Impedance continuously to
select the input impedance of the oscilloscope. If you select "50 Ω", the channel
status label at the bottom of the screen will display the icon
.
Note:
After the oscilloscope recognizes the probe automatically, the input impedance
will also be auto recognized. You do not have to set it manually.
The setting of the input impedance will affect the ranges of channel vertical
scale and offset.