RIGOL Chapter 6 Operations and Measurements
6-14 MSO7000/DS7000 User's Guide
If Source A (or Source B) selects CH3, press More Thre.CH3 to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
If Source A (or Source B) selects CH4, press More Thre.CH4 to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
Press Offset, and then rotate the multifunction knob to adjust the vertical
offset of the operation results. You can also directly rotate Wave Vertical
POSITION or enable the touch screen to adjust the vertical offset.
Press Size, and then rotate the multifunction knob to select the waveform
size. Press down the knob to select it. You can also press Size continuously or
enable the touch screen to select it. The available sizes include "Small",
"Medium", and "Large". Also, you can rotate Wave Vertical
SCALE to
adjust the waveform size.
Press AutoScale to adjust the vertical scale and offset of the operation results
to the optimal value according to the current configuration, so as to better
observe.
Press More Sensitivity to set the sensitivity of the digital signal converted
from the analog signal on the source. For setting methods, refer to descriptions
in "Parameter Setting Method".
Press More Label to set the label for math operation results. For details,
refer to descriptions in "Math Operation Label".
Intg
Calculates the integral of the selected source. For example, you can use integral to
measure the area under a waveform or the pulse energy.
Press Math Math1 Operator to select "Intg":
Press Operation continuously to enable or disable the "Intg" operation
function.
Press SourceA, and rotate the multifunction knob to select the channel for
SourceA. Then press down the knob to select the channel. You can also press
SourceA continuously or enable the touch screen to select the channel. The
available channels for SourceA are CH1, CH2, CH3, and CH4.
Note: Only the channels that have been enabled currently can be selected.
Press Offset to set the vertical offset of the operation results. For setting
methods, refer to the descriptions in "Parameter Setting Method". You can
also directly rotate Wave Vertical
POSITION or enable the touch screen
to adjust the vertical offset.
Press Scale to set the vertical scale of the operation results. For setting methods,
refer to the descriptions in "Parameter Setting Method". You can also
directly rotate Wave Vertical
SCALE or enable the touch screen to adjust