Chapter 6 Operations and Measurements RIGOL
MSO7000/DS7000 User's Guide 6-41
include AY, BY, and â–³Y, and they are all expressed in percentage.
This oscilloscope allows you to set the reference position for the ratio
according to your needs. After adjusting Y Cursor to a desired position,
press Set Reference to take the current cursor position as the
reference. The current positions of Cursor A and Cursor B are
respectively defined as "0%" and "100%". Meanwhile, two cursors (in
blue) that cannot be moved are displayed on the screen as the ratio
reference position (the ratio position at "0%" and "100%" are
respectively marked by a solid line and a dotted line). Before you set
manually, the oscilloscope adopts the default ratio reference position.
6. Set LA Type
When "LA" is selected under Source, you need to set the LA display type. Press
More and then press La Type continuously to select "HEX" or "BIN".
7. Measurement Example
Measure the period of a sine wave by using the manual cursor measurement
and auto measurement respectively. The measurement results are both 1 ms, as
shown in Figure 6-3.
Figure 6-3 Manual Cursor Measurement Example
Track Mode
In this mode, you can adjust the two pairs of cursors (Cursor A and Cursor B) to
measure the X and Y values on two different sources respectively. The points being
measured on Cursor A and Cursor B are marked by
and respectively.
When the cursors are moved horizontally/vertically, the markers will position on the
waveform automatically. When the waveform is expanded or compressed
Auto
Measurement
Cursor
Measurement