Loading ...
Loading ...
Loading ...
SIGLENT
SDS2000X User Manual 57
If the signal you are probing is noisy, you can setup the oscilloscope to reduce the noise in
the trigger path and on the displayed waveform. First, stabilize the displayed waveform by
removing the noise from the trigger path. Second, reduce the noise on the displayed
waveform.
1. Connect a signal to the oscilloscope and obtain a stable display.
2. Remove the noise from the trigger path by setting trigger coupling to LF Reject, HF
Reject or turning on Noise Reject.
3. Set the Acquisition option to Average to reduce noise on the displayed waveform.
Loading ...
Loading ...
Loading ...