Siglent SDS-2000X-LA MSO Function Software for SDS2000X Series Oscilloscopes

Manual - Page 164

For SDS-2000X-LA.

PDF File Manual, 250 pages, Read Online | Download pdf file

SDS-2000X-LA photo
Loading ...
Loading ...
Loading ...
SIGLENT
134 SDS2000X User Manual
To Make Cursor Measurements
1. Press the Cursors button on the front panel to enter the CURSOR function menu.
2. Press the Mode softkey and set the option to On.
3. Press the Source softkey, and then use the Universal Knob to select the desired
source. Only analog channels (CH1~4), math waveforms and reference waveforms
(REFA, REFB, REFC, REFD) that are displayed are available for cursors.
4. To make cursor measurements:
To measure the horizontal time, use the Universal Knob to move the X1 and X2
cursors to desired place. If necessary, set the cursor type to X1-X2 and move the
X1 and X2 cursors together.
To measure vertical voltage or current, use the Universal Knob to move the Y1
and Y2 cursors to the desired position. If necessary, set the cursor type to “Y1-Y2”
and move the Y1 and Y2 cursors together.
To adjust the transparency of the cursors message box, press the
Display/Persist button and go to the second page, press the Transparence (20%
to 80%) softkey and then turn the Universal Knob to adjust the transparence to
the desired value.
Cursor examples:
1. Use cursors to measure pulse width:
Figure 73: Measure Pulse Width