Siglent SDS-2000X-LA MSO Function Software for SDS2000X Series Oscilloscopes

Manual - Page 86

For SDS-2000X-LA.

PDF File Manual, 250 pages, Read Online | Download pdf file

SDS-2000X-LA photo
Loading ...
Loading ...
Loading ...
SIGLENT
56 SDS2000X User Manual
Noise Rejection
Noise Reject adds additional hysteresis to the trigger circuitry. By increasing the trigger
hysteresis band, you reduce the possibility of triggering on noise. However, this also
decreases the trigger sensitivity so that a slightly larger signal is required to trigger the
oscilloscope.
Press the Setup button on the front panel, and then press the Noise Reject softkey
continually to set the option to On or Off to turn on or off the noise rejection function.
Figure 22: Turn off Noise Reject
Figure 23: Turn on Noise Reject