Siglent SDS-2000X-LA MSO Function Software for SDS2000X Series Oscilloscopes

Manual - Page 130

For SDS-2000X-LA.

PDF File Manual, 250 pages, Read Online | Download pdf file

SDS-2000X-LA photo
Loading ...
Loading ...
Loading ...
SIGLENT
100 SDS2000X User Manual
CAN Trigger and Serial Decode
The following section describes trigger and decoding for the CAN bus and is
organized as follows: Setup for CAN Signals, CAN Triggering and CAN Serial
Decode”.
Setup for CAN Signals
1. Press the Decode key to enter the DECODE function menu.
2. Press the Decode softkey and select the desired slot (Decode1 or Decode2).
3. Press Protocol softkey and then select CAN by turning Universal Knob.
4. Press Signal softkey to enter the SIGNAL menu as Figure 52 shows.
Figure 52: CAN SIGNAL Menu
5. Set CAN-H:
a. Press CAN-H softkey to select the channel that is connected to the CAN-H
signal.
b. Press first Threshold key to set the CAN-H signal threshold voltage level by the
Universal Knob. The threshold voltage level is for decoding, and it will be
regarded as the trigger voltage level when set the trigger type to serial.
6. Set CAN-L:
a. Press CAN-L to select the channel that is connected to the CAN-L signal.
b. Press second Threshold key to set the CAN-L signal threshold voltage level by
the Universal Knob. The threshold voltage level is for decoding, and it will be
regarded as the trigger voltage level when set the trigger type to serial.
7. Press the Configure softkey to enter the BUS CONFIG menu.
8. Press Baud to set baud rate by the Universal Knob.
The baud rate can be set as a predefined value (from 5kb/s to 1Mb/s) or custom
value (from 1b/s to 1Mb/s).