compensated or poorly compensated probes can introduce measurement inaccuracy
or errors. The compensation procedure is as follows:
1. Perform step 1, 2, 3, and 4 in
Function Inspection
.
2. Check the displayed waveforms and compare them with waveforms shown in the
figure below.
Figure 4.11 Probe Compensation
3. Use the probe compensation adjustment tool provided in the accessories to adjust
the low-frequency compensation adjustment hole on the probe until the displayed
waveform is consistent with the "Perfectly compensated" waveform shown in the
figure above.
4.4 Product Overview
The following content provides a description of the front and rear panels as well as
the user interface of the DHO4000 series by taking DHO4804 as an example.