DHO4000 series user manual - Page 194

For DHO4204.

PDF File Manual, 302 pages, Read Online | Download pdf file

DHO4204 photo
Loading ...
Loading ...
Loading ...
You can click or tap the "Counter" label in the "Result" bar and select Reset Stat,
Setting, or Remove in the displayed window.
11.2.1 Measurement Settings
After the frequency counter is enabled, click or tap the "Counter" label in the "Result"
bar at the right section of the screen and a window is displayed. Click or tap Setting
in the window to enter the counter setting menu, as shown in the figure below.
Figure 11.2 Frequency Counter Setting Menu
Select the Measurement Source
Click or tap the drop-down button of Source to select the desired source. Analog
channels (CH1-CH4) and EXT can be selected as the source of the frequency counter.
Set Resolution
For Period and Frequency measurements, you need to set the readout resolution.
Click or tap the input field of
Resolution to set the resolution by using the pop-up
numeric keypad. You can also use the corresponding multipurpose knob to set the
value. The range of resolution is from 3 bits to 6 bits. By default, it is 4 bits.
Digital Voltmeter (DVM) and Frequency Counter
DHO4000 User Guide
178
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.