After a trigger type is selected, the current trigger setting information (including
trigger type, trigger source, and trigger level) is displayed in the trigger information
label at the top of the screen, as shown in the figure below. The information will
change based on the trigger settings.
Source Selection
Click or tap the drop-down button of Source to select CH1-CH4. For details, refer to
descriptions in
Trigger Source
The selected trigger source is indicated in the trigger
information label at the top of the screen.
Only when you select the channel that has signal inputs as the trigger source, can you
obtain a stable trigger.
Polarity
In the Polarity item, select the desired polarity: positive polarity ( ) or negative
polarity ( ).
Level Selection and Setting
After the trigger condition setting is completed, you need to adjust the trigger level
to correctly trigger the signal and obtain a stable waveform.
Click or tap the Level A/Level B input field to set the level A/level B with the pop-up
numeric keypad. You can also use the trigger level knob or use the corresponding
multipurpose knob to adjust level A/level B. When Linkage is ticked, the upper limit
and lower limit values change at the same time. The difference between upper and
lower limit remains unchanged. For details, refer to descriptions in
Trigger Level
. The
current trigger level is displayed in the trigger information label at the top of the
screen.
TIP
Press the trigger level knob to switch between three modes: "only adjust Level A", "only adjust
Level B", and "adjust Level A and Level B at the same time".
Trigger Condition
Click or tap the drop-down button of
When to select the desired trigger condition.
• Sync: triggers on the specified sync type. After this trigger condition is selected,
click or tap the drop-down button of Sync to select the desired sync type: Data
Sync, Cmd/Status Sync, or All Sync.
Triggering the Oscilloscope