Operation Steps:
1. Press Meter to enter multimeter mode, press F1 to choose DCV, ACV
measurement.
2. Insert the red probe to the V.Ω.C banana jack input and the black probe to the
COM. Connect the other end of probes to the power or load to be measured.
3. Turn on /off the relative according to the real demand.
4. Choose Manual or Auto according to the real demand.
5. Read voltage value.
Making Resistance Measurement
Figure 3-4 Resistance Measurement
Operation steps:
1. Press Meter to enter multimeter mode, press F1 to choose Res.
measurement.
2. Insert the red probe to the V.Ω.C banana jack input and the black probe to the
COM. Connect the other end of probes to the power or load to be measured.
3. Turn on /off the relative according to the real demand.
4. Choose Manual or Auto according to the real demand.
5. Read resistance value.
Notes:
When measuring resistant, please make sure that the circuit is power off and the
capacitance is discharged to avoid damage to the SHS800.