SHS800 Series 39
Self Calibration
Self Calibration is operated to calibrate the relative data of SHS800 to decrease the
mistake during the measure. If the operating temperature changes by or more than 5° C
or the instrument runs more than thirty minutes, you should do the self calibration. When
you do the self calibration, you should cut off all the probes and leads. Then press
User button to choose Do self cal to show the self calibration menu, and do self
calibration program according to the prompts on the screen.
Figeure 2-62 Self Calibration
Print Setting
The SHS800 supports PictBridge compatible printers. You can connect the side USB
Device of the SHS800 to the USB Device of the PictBridge compatible printer through
USB cable. After setting the print settings, press Save/Recall button for 4 seconds
to complete the print operation.
Figure 2-63 Print Setting Menu 1
Table 2-40 Print Setup Function Menu 1
Option Setting Introduction
Ink Saver
On
Off
Print the screen image on a white
background when you select On.
Layout Portrait/Landscape The output direction of the printer.
Paper Size
Set the type of the
paper.
Displays settings available on your
PictBridge compatible printer.