66 SDS1000CFL User Manual
Table 2-40 All Measurement function menu:
Option Setting Instruction
Source
CH1
CH2
CH3
CH4
Select input signal channel.
Voltage
On
Off
Turn on the all measurement function to
measure voltage parameters.
Turn of the all measurement function to
measure voltage parameters.
Time
On
Off
Turn on the all measurement function to
measure Time parameters.
Turn of the all measurement function to
measure Time parameters.
Return Return to the “All Measure main menu”.
Table 2-41 Introduction of the Measure Type:
Measure Type Introduction
Vmax
The most positive peak voltage measured over the
entire waveform.
Vmin
The most negative peak voltage measured over the
entire waveform.
Vpp
Measures the absolute difference between the
maximum and minimum peaks of the entire waveform.
Vtop
Measures the highest voltage over the entire
waveform.
Vbase
Measures the lowest voltage over the entire waveform.
Vamp
Voltage between Vhig and Vlow of a waveform.
Vavg
The arithmetic mean over the first cycle in the
waveform.
Mean
The arithmetic mean over the entire waveform.
Crms
The true Root Mean Square voltage over the first cycle
in the waveform.
Vrms
The true Root Mean Square voltage over the entire
waveform.
ROVShoot
Defined as (Vmax-Vhig)/Vamp after the waveform
rising.
FOVShoot
Defined as (Vmin-Vlow)/Vamp after the waveform
falling.
RPREshoot
Defined as (Vmin-Vlow)/Vamp before the
waveform rising.
FPREshoot
Defined as (Vmax-Vhig)/Vamp before the
waveform falling.
Rise Time
Rise Time Measures the time between 10% and 90%
of the first rising edge of the waveform.