SDS1000CFL User Manual - Page 56

For SDS1000CFL.

PDF File Manual, 134 pages, Download pdf file

Loading ...
Loading ...
Loading ...
SIGLENT
46 SDS1000CFL User Manual
When
Select slope trigger condition.
Next Page
Page 1/2
Enter the second page of the alternative
trigger.
Table 2-28 Set trigger mode to slope trigger function menu 2
Option Setting Instruction
Time
Set
time
Turn the “Universal” knob to set the slope
time. Time setup range is 20ns-10s.
Vertical
Select the trigger level that can be adjusted
by LEVEL” knob. You can adjust “LEVEL
A”, “LEVEL B” or adjust them at the same
time.
Set up Enter “Trigger setup menu”(see table 2-14).
Next Page Page 2/2 Return to the first page of “Alternative trigger
menu”.
Operate Instruction:
Observe two irrelative channel signals, follow these steps
1. Input two irrelative signals to channel 1and channel 2.
2. Press the AUTO button.
3. Press the TRIG MENU button to enter “trigger menu”.
4. Press the “Type” option button to select “Alternative”.
5. Press the “Channels” option button to select “CH1-CH2”
6. Press the “Source” option button to select “CH1”.
7. Press the CH1 button and turn the “S/div” knob to optimize waveform
display.
8. Press “Mode” option button to select “Edge”, “Pulse”, “Slope” or “Video”.
9. Set the trigger according to trigger edge.
10. Press the “Source” option button to select “CH2”.
11. Press the CH2 button and turn the “S/div” knob to optimize waveform
display.
12. Repeat steps 8 and 9.