118 SDS1000CFL User Manual
offset range 102mV-5V: ±40V)
Vertical Resolution 8 bit
Channels 2
Analog
Bandwidth
300MHz(SDS1304CFL, SDS1302CFL)
200MHz(SDS1204CFL, SDS1202CFL)
100MHz(SDS1104CFL, SDS1102CFL)
70MHz(SDS1074CFL, SDS1072CFL)
Single-shot
Bandwidth
300MHz(SDS1304CFL, SDS1302CFL)
200MHz(SDS1204CFL, SDS1202CFL)
100MHz(SDS1104CFL, SDS1102CFL)
70MHz(SDS1074CFL, SDS1072CFL)
BW Flatness at BNC
input
DC -10% of rated BW: +/- 1dB
10% - 50% of rated BW: +/- 2dB
50% - 100% of rated BW: + 2dB/-3dB
Lower frequency limit
(AC -3dB)
≤10Hz(at input BNC)
Noise: Pk-Pk for 3K
record
≤0.6 Div for average of 10 Pk-Pk readings, Fixed
gain settings
≤0.7 Div for average of 10 Pk-Pk readings,
Variable gain settings
SFDR including
harmonics
(measured with FFT)
>=35dB
DC Gain Accuracy
± 3.0%: 5mv/div to 5V/div in Fixed Gain
Ranges
<±4.0%: 2mv/div Variable Gain Ranges
DC Measurement
Accuracy:
All Gain settings
100mv/div
± [3%* |reading|+|offset| +1% *of |offset|
+0.2div+2mv]
DC Measurement
Accuracy:
All Gain settings
100mv/div
± [3%* |reading|+|offset| +1%* of |offset|
+0.2div+100mv]
Rise time
<1.2ns ( SDS1304CFL, SDS1302CFL )
<1.8ns ( SDS1204CFL, SDS1202CFL )
<3.5ns ( SDS1104CFL, SDS1102CFL )
<5.0ns ( SDS1074CFL, SDS1072CFL )
Overshoot, Typical
(using 500ps pulse)
<10% with probe or BNC input w/ 50 Ohm feed
thru
Ch to Ch Skew (both
channels in same
V/div setting)
<1ns: SDS1304CFL,SDS1302CFL
SDS1204CFL,SDS1202CFL
SDS1104CFL,SDS1102CFL
<2ns: SDS1074CFL,SDS1072CFL
(Equivalent to 2 minor divisions in smallest t/div)
Math operation +, -, *, /, FFT
FFT
Window mode: Hanning, Hamming, Blackman,
Rectangular
Loading ...