Siglent SDS1000CFL

SDS1000CFL User Manual - Page 52

For SDS1000CFL.

PDF File Manual, 134 pages, Download pdf file

Loading ...
Loading ...
Loading ...
SIGLENT
42 SDS1000CFL User Manual
Table 2-21Slope trigger function menu 2
Option Setting Instruction
Type Slope
Trigger on positive slope of negative slope
according to setup time of the oscilloscope.
Vertical
Select the trigger level that can be adjusted by
“LEVEL” knob. You can adjust “LEVEL A”,
“LEVEL B” or adjust them at the same time.
Auto
Use this mode to let the acquisition free-run in
the absence of a valid trigger; This mode
allows an untriggered, scanning waveform at
100 ms/div or slower time base settings.
Normal
Use this mode when you want to see only valid
triggered waveforms; when you use this mode,
the oscilloscope does not display a waveform
until after the first trigger.
Mode
Single
When you want the oscilloscope to acquire a
single waveform, press the “SINGLE” button.
Set up Enter “Trigger setup menu” (See table 2-14).
Next Page
Page 2/2 Return the first page of slope trigger.
Picture 2.8-7
Operate Instruction:
Follow the next steps if you select “Slope trigger”:
1. Input a signal to one channel.
2. Press the “AUTO” button.
Loading ...
Loading ...
Loading ...