SIGLENT
SDS1000CFL User Manual 117
Appendix A: Specifications
All specification applies to 10X probe and All the SDS1000CFL Series Digital
Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must first
meet the following conditions:
● The oscilloscope must have been operating continuously for thirty minutes
within the specified operating temperature.
● You must perform the Do Self Cal operation, accessible through the Utility menu,
if the operating temperature changes by more than 5° C.
● The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Specifications
Inputs
Input Coupling AC, DC, GND
Input Impedance
1MΩ±2% || 18Pf±3Pf,
50Ω+/-2%(SDS1304CFL/SDS1302CFL,
SDS1204CFL/SDS1202CFL contain this
function)
Maximum input
voltage
400V (DC+AC PK-PK,1MΩ input impedance,
X10),CAT I
Ch to Ch Isolation
(Both channels in
same V/div setting)
>100:1 at 150MHz
(SDS1304CFL,SDS1302CFL)
>100:1 at 100MHz
(SDS1204CFL,SDS1202CFL)
>100:1 at 50MHz (SDS1104CFL,SDS1102CFL)
>100:1 at 35MHz (SDS1074CFL,SDS1072CFL)
Probe attenuator 1X,10X
Probe attenuator
Factors Set
1X,5X,10X,50X,100X,500X,1000X
Vertical System
Vertical Sensitivity 2mV/div -5V/div(1-2-5 order)
Channel voltage
2mV-100mV: ±800MV