Operate Instruction:
1. Set up Type
1) Press the “TRIG MENU” button to display “Trigger” menu.
2) Press the “Type” option button to select “Edge”.
2. Set up Source
According to input signal, press the “Source” option button to
select “CH1”, “CH2”,”CH3”,”CH4”,“EXT”, “EXT/5”or “AC Line”.
3. Set up Slope
Press the “Slope” option button to select ” , ” or .
4. Set up Trigger mode
Press the “Trigger mode” option button to select “Auto”, “Normal”,
“Single”.
Auto: The waveform refresh at a high speed whether the trigger
condition is satisfied or not.
Normal: The waveform refresh when the trigger condition is satisfied
and waits for next trigger event occurring when the trigger condition is
not satisfied.
Single: The oscilloscope acquire a waveform when the trigger
condition is satisfied and then stops.
5. Set up Trigger coupling
1) Press the “Set up” button to enter the “Trigger Setup Menu”.
2) Press the “Coupling” option button to select “DC”, “AC”, “HF Reject”
or “LF Reject”.