SDS1000CFL User Manual 57
2.10.1 X-Y Format
Use the XY format to analyze phase differences, such as those represented by
Lissajous patterns. The format plots the voltage on channel 1 against the voltage
on channel 2, or plots the voltage on channel 3 against the voltage on channel 4,
where channel 1(or channel 3) is the horizontal axis and channel 2(or channel 4)
is the vertical axis. The oscilloscope uses the untriggered Sample acquisition
mode and displays data as dots.
XY mode has a breakthrough that trad oscilloscopes restrict sampling rate at
1MSa/s,
support 25KSa/s~250MSa/s (1-2.5-5 sequence) adjusted ;
Operation steps :
● Channel 1(or channel 3) “Volt/div”and vertical “POSITION” set up the
horizontal scale and position.
● Channel 2(or channel 4) “Volt/div”and vertical “POSITION” set up the
horizontal scale and position.
●Turn the “S/div” knob to adjust the sampling rate
●The following functions are forbidden at XY display form:
●Benchmark wave form and wave mathematic
●Cursor
●Auto(resets display format to YT)
●Trigger Control
●Horizontal Position Knob
●Vector Display Type
●Scan Display
Note:The oscilloscope can capture a waveform in normal YT
mode at any sampling rate. You can view the same waveform
in XY mode.
To do so, stop the acquisition and change th
e display format