SDS1000CFL User Manual 35
Auto
Use this mode to let the acquisition free-run in the
absence of a valid trigger; This mode allows an
untriggered, scanning waveform at 100 ms/div or
slower time base settings.
Normal
Use this mode when you want to see only valid
triggered waveforms; when you use this mode, the
oscilloscope does not display a waveform until after
the first trigger.
Mode
Single
When you want the oscilloscope to acquire a single
waveform, press the “SINGLE ”button.
Set up
Enter the “Trigger Setup Menu” (See table 2-14).
Table 2-15 Trigger Setup function menu
Option Setting Explain
DC Passes all components of the signal
AC
Blocks DC components attenuates signals
below 25 Hz.
HF Reject
Attenuates the high-frequency components
above 150 kHz.
Coupling
LF Reject
Blocks the DC component , attenuates the
low-frequency components below 8 kHz.
Holdoff
Using the “universal” knob to adjust holdoff
time(sec)the holdoff value is displayed.
Holdoff
Reset
Reset holdoff time to 100ns.
Return Return the first page of “Trigger main menu”.