Siglent SDS1000CFL

SDS1000CFL User Manual - Page 49

For SDS1000CFL.

PDF File Manual, 134 pages, Download pdf file

Loading ...
Loading ...
Loading ...
SIGLENT
SDS1000CFL User Manual 39
Video TriggerTrigger on fields or lines of standard video signals.
Table 2-18 Functional Manu of Video Trigger 1:
Option Setting Instruction
Type Video
When you select the video type, put the
couple set to the AC, then you could
trigger the NTSC,PAL and SECAM video
signal.
CH1CH2
CH3CH4
Select the input source to be the trigger
signal.
Source
EXT
EXT/5
Ext and Ext/5 use the signal applied to
the EXT TRIG connector as the source.
(Normal)
Normal triggers on the negative edge of
the sync pulse.
Polarity
(Inverted)
Inverted triggers on the positive edge of
the sync pulse.
Sync
Line Num
All lines
Odd field
Even field
Select appropriate video sync.
Next Page Page 1/2
Enter the second page of “Video trigger
menu”.
Table 2-19 Functional Manu of Video Trigger 2:
Option Setting Instruction
Type Video
When you select the video type, put
the couple set to the AC, then you
could trigger the NTSC,PAL and
SECAM video signal.
Standard
NTSC
Pal/Secam
Select the video standard for sync
and line number count.
Auto
Use this mode to let the acquisition
free-run in the absence of a valid
trigger; This mode allows an
untriggered, scanning waveform at
100 ms/div or slower time base
settings.
Normal
Use this mode when you want to see
only valid triggered waveforms; when
you use this mode, the oscilloscope
does not display a waveform until
after the first trigger.
Mode
Single
when you want the oscilloscope to
acquire a single waveform, press the
“SINGLE ”button.
Set up Enter the “Trigger setup menu”.
Next Page Page 2/2
Return the first page of “Video
Trigger menu”.