Average: The oscilloscope acquires several waveforms, averages them, and
displays the resulting waveform.
Advantage: You can use this mode to reduce random noise.
Picture 2.9-4 Average mode
■ Equivalent Time Sampling:
The equivalent time sampling mode can achieve up to 20 ps of horizontal
resolution (equivalent to 50GSa/s). This mode is good for observing repetitive
waveforms.
Real Time Sampling: The scope has the highest Real-time sampling rate
up to 2GSa/s.
■ “RUN/STOP” Button: Press the RUN/STOP button when you want the
oscilloscope to acquire waveforms continuously. Press the button again to stop
the acquisition.
■ “SINGLE” Button: Press the SINGLE button to acquire a single waveform.
Each time you push the SINGLE button, the oscilloscope begins to acquire
another waveform. After the oscilloscope detects a trigger it completes the
acquisition and stops.
When you push the RUN/STOP or SINGLE buttons to start an acquisition, the
oscilloscope goes through the following steps:
1. Acquire enough data to fill the portion of the waveform record to the left of
the trigger point. This is also called the pre-trigger.