CB-S / CB-S-UL (E7) 06/2020 page 92/119
Do not open the chamber doors until the interior temperature has dropped to 37 °C / 98.6 °F.
After premature termination the chamber reverts to its standard operational status.
The CO
2
actual value display shows because the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the chamber will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing the OK button.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F
Start up the chamber (chap. 6).
23.5.2 Opening the outer door
For safety reasons, the sterilization cycle is automatically aborted if you open the outer chamber door.
Do not open the glass door and close the outer door immediately.
After premature termination caused by opening the outer door the chamber reverts to its standard
operational status.
The CO
2
actual value display shows because the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the chamber will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing the OK button.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F
Start up the chamber (chap. 6).
23.5.3 Turning off the chamber
For safety reasons, the sterilization cycle is aborted if the chamber is turned off or if there is a power fail-
ure.
Normally you should not terminate sterilization in this way.
Do not open the chamber doors until the interior temperature has dropped to 37 °C / 98.6 °F.
After turning on again the chamber at the main power switch, it will revert to its standard operational
status.
The CO
2
actual value display shows because the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the chamber will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing the OK button.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO
2
sensor.