CB-S / CB-S-UL (E7) 06/2020 page 25/119
3.2 Guidelines for safe lifting and transportation
After operation, please observe the guidelines for temporarily decommissioning the chamber (chap. 25.2).
Empty the water pan before moving the incubator. In case of any spilling of the contents, shut down the
incubator and dry it out carefully and completely
CAUTION
Risk of injury and damages by lifting heavy loads and by sliding or tilting of the
chamber due to improper transportation.
Injuries, damage to the chamber.
Transport the chamber in its original packaging only.
For moving or shipping, secure the chamber with transport straps.
Do NOT lift or transport the chamber using the door handle or the door.
Lift the chamber at the four lower corners with the aid of 4 people and place it on a
rolling pallet.
Move the chamber to the desir
ed location and lift it from the rolling pallet with the aid of
four people.
Permissible ambient temperature range for transport: -10 °C / 14°F to +60 °C / 140°F.
You can order transport packing and rolling pallets for transportation purposes from BINDER Service.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 25.2).
Permissible ambient temperature range for storage: -10 °C / 14°F to +60 °C / 140°F.
Permissible ambient humidity: max. 70% r.h., non-condensing
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form in the inner chamber, on the housing or in the sensor compartment of the CO
2
measure-
ment. Before start-up, wait at least one hour until the CO
2
incubator has attained ambient temperature
and is completely dry.
3.4 Location of installation and ambient conditions
Notes on the location of installation
Set up the chamber on a flat, even surface, free from vibration and in a well-ventilated, dry location. The
chambers are designed for setting up inside a building (indoor use).
Freestanding chambers are suitable for installation on tables or on the optionally available stand. Note:
The site of installation must be capable of supporting the chamber’s weight (see technical data, chap.
26.4).
Align the chamber using a spirit level to ensure even covering of the cell-cultures with the medium. For
this purpose, manually adjust the four incubator feet.
In order to avoid contamination, never place the chamber directly on the floor.