CB-S / CB-S-UL (E7) 06/2020 page 89/119
23.3.2 Performance during sterilization
Normal display during sterilization (example during heating up phase)
Sterilization
The information messageSterilizationis shown alternating with the indication “Temp. / CO2”.
The CO
2
actual value display shows “ ” because the CO
2
sensor is disconnected.
The zero-voltage relay alarm output and the buzzer are not activated.
When starting a hot-air sterilization, CO
2
control deactivates automatically.
The safety controller settings are inactive during sterilization. They become functional again
following termination of the sterilization and/or restarting the chamber at the main power
switch.
Glass door and inner chamber become hot during sterilization.
CAUTION
Danger of burning by touching hot chamber parts during a sterilization.
Burns.
Do NOT touch the glass door, glass door, inner surfaces, and door gaskets during a
sterilization.
Opening the outer door leads to prematurely terminating the running sterilization cycle (chap. 23.5). If this
happens during the heating up or holding phase (during 6 hours after starting the sterilization cycle), steri-
lization will be ineffective (chap. 23.4.1).
NOTICE
Danger of interrupting the temperature reaction time by opening the outer door
during the heating up or holding phase.
Ineffective sterilization.
Do NOT open the chamber doors during the heating up or holding phase of the sterili-
zation cycle
23.3.3 Completing the entire sterilization cycle
After 6 hours the information message “Sterilization success” indicates the successful sterilization. The
cooling down phase starts.
The entire sterilization cycle ends after a total of approx. 12 hours.
When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F
Start up the chamber (chap. 6)
The chamber is now ready to operate.