CB-S / CB-S-UL (E7) 06/2020 page 63/119
15.4.8 Alarms referring to CO
2
sensor failure
If CO
2
control has been deactivated (chap. 9.2), no alarms of the CO
2
gas system will be
emitted.
The alarms occur immediately.
Failure of CO
2
sensor
Actual CO
2
value display shows – ”
Alarm message CO2 sensor defective”, “collective alarmicon
Actions:
Turn off the chamber.
Remove the sensor (chap. 4.2), replace it if necessary.
If necessary, clean, disinfect and sterilize the chamber before restarting it
Please contact BINDER Service.
CO
2
sensor not connected
Actual CO
2
value display shows – ”
Actions:
Turn off the chamber.
Connect the sensor (chap. 4.2.1)
15.4.9 Alarm on ineffective sterilization
The sterilization cycle was terminated prematurely after less than 6 hours (chap. 23.4.1)
Alarm message Sterilization failed”, “collective alarmicon
Actions:
Repeat the sterilization, if necessary (chap. 23.3).