12.4 Setting the current time
Required access level: “Admin”. Following start-up of the chamber (chap. 6.3), it is “User”.
Path: Normal display Settings Chamber Time
Press the OK button to enable the setting.
Setting the time: hours
The current setting flashes. Enter the current hour with the arrow but-
tons.
Confirm the entry with the OK button.
Time
Setting the time: minutes
The current setting flashes. Enter the current minutes with the arrow
buttons.
Confirm the entry with the OK button.
Time
With the Back button you can go back to the “Chamber” submenu and, repeatedly pressing it, to Nor-
mal display.
12.5 Function “Language selection at restart”
If the function “Language selection at restart” is activated, menu language, date, time, and temperature
unit are checked with every startup of the chamber. At this occasion it is also possible to modify them with
“User” access level.
Required access level: “Admin”.
Path: Normal display Settings Chamber Language selection at restart
Press the OK button to enable the setting.
Function “Language selection at restart”
The current setting flashes. Use the arrow buttons to select between
ON and OFF.
Confirm the setting with the OK button.
Language sel. at restar
With the arrow-down button you can now change to the next parameter (chamber address).
With the Back button you can go back to the “Chambersubmenu and, repeatedly pressing it, to Nor-
mal display.