1.14 Measures to prevent accidents
The manufacturer took the following measures to prevent dangers:
Indications on the type plate
See operating manual chap. 1.6.
Operating manual
An operating manual is available for each chamber.
Overtemperature monitoring
The chamber is equipped with a temperature display, which can be read from outside.
The chamber is equipped with an additional safety controller (temperature safety device class 3.1 acc.
to DIN 12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding.
Safety, measurement, and control equipment
The safety, measuring, and control equipment is easily accessible.
Electrostatic charge
The interior parts are grounded.
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electri-
cal equipment (e.g. power cables). The machine is equipped with no permanent magnets. If persons
with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field source to
implant) of 30 cm, an influence of these implants can be excluded with high probability.
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
Floors
See operating manual chap. 3.4 for correct installation
Cleaning
See operating manual chap. 22.1.