CB-S / CB-S-UL (E7) 06/2020 page 53/119
13.3 Setting the CO
2
tolerance range
Path: Normal display Settings Various CO2 range
Press the OK button to enable the setting.
Setting the CO
2
tolerance range
The current setting flashes. Enter the desired CO
2
range with the ar-
row buttons. Entry range: 1.0 vol.-% up to 10.0 vol.-%
Confirm the entry with the OK button.
CO2 range
With the arrow-up button you can go back to the temperature range setting
With the arrow-down button you can now change to the door open alarm delay setting.
With the Back button you can go back to the “Varioussubmenu and, repeatedly pressing it, to Normal
display.
13.4 Setting the delay time for door open alarm
Path: Normal display Settings Various Door alarm delay
(min)
Press the OK button to enable the setting.
Setting the door open alarm delay.
The current setting flashes. Use the arrow buttons to enter the de-
sired time after which the door alarm shall be triggered. Entry range: 1
up to 120 minutes.
Confirm the entry with the OK button.
Door alarm delay (min)
With the arrow-up button you can go back to the CO
2
tolerance range setting
With the arrow-down button you can now change to the humidity control setting.
With the Back button you can go back to the “Varioussubmenu and, repeatedly pressing it, to Normal
display.