CB-S / CB-S-UL (E7) 06/2020 page 61/119
Actions:
Check that you have set the pressure on the pressure reducer at 2.0 bar / 29 psi above the ambient
air pressure, and that all the valves are open for the gas supply.
Where the CO
2
supply is a pressurized gas cylinder, check that the cylinder still contains sufficient
CO
2
. If necessary, replace the gas cylinder. Observe the precautions when handling gases and the
correct outlet pressure (chap. 4.4).
When using central CO
2
supply, check whether the primary pressure is high enough.
Check that the gas tube has no damage, kinks, blockages or soiling.
Check when the gas filter was last replaced. Replace the gas filter every year to avoid it blocking. A
qualified service engineer should replace the gas filter.
If these points do not reveal the source of the fault, it may be that the chamber is faulty. Please contact
BINDER Service.
The outlet pressure of the gas cylinder must be 2.0 bar / 29 psi above the ambient pressure.
NOTICE
Danger of damage by excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the chamber.
The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.
Before connecting, check the outlet pressure on the pressure reducer of the cylinder.
Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure.
The recovery times of the gas concentrations inside the chamber following the door being opened, which
are indicated in the technical data (chap. 26.4), refer to a connection pressure of 2.0 bar / 29 psi. De-
creasing supply pressure down to the alarm point of 0.3 bar / 4.4 psi results in longer recovery times.
Check the pressure displays of your gas supply. If very short recovery times are required or the door is
opened frequently, replace the gas cylinders promptly when the pressure decreases below 2.0 bar / 29
psi.
15.4.6 Power failure alarm
No alarm message (display is off)
Audible alarm (buzzer): immediately
Switching the zero-voltage relay alarm output.
WARNING: If customer should use a chamber running in non-supervised continuous opera-
tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to
split such specimen or samples and store them in at least two chambers, if this is feasible.