Binder CB-S 260 CO₂ incubator with hot air sterilization, series CB – 220-liter capacity

User Manual - Page 42

For CB-S 260.

PDF File Manual, 119 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB-S / CB-S-UL (E7) 06/2020 page 42/119
9. Setting special controller functions
In the “Functions on/off” menu you can define the switching state of the special controller functions.
Required access level: “User”.
Path: Normal display Setpoints Functions on/off
The functions are displayed from left to right.
Example: Function 1 “Humidity off” activated = 1000. Function 1 “Humidity off” deactivated = 0000.
Submenu “Functions on/off”.
This view shows the switching states four functions.
“1” = Function activated
“0” = Function deactivated
Functions on/off
Press the OK button to access the first individual function. With the arrow-down button you can pro-
ceed to the subsequent functions.
Function 1 Idle mode”: Change to Idle mode, chap. 9.1
Function 2 CO2 off”: Deactivating CO
2
control, chap. 9.2
Functions 3 and 4 have no function with this chamber type.
Press the OK button to enable the setting of the desired function and select the function’s switching state
“1” (function activated) or “0” (function deactivated).
With the Back button you can go back to the “Setpoints” submenu and, repeatedly pressing it, to Nor-
mal display.
In Normal display the activated functions are indicated by an icon showing the number of the respective
function.
Example:
Normal display with activated function 1 “Idle mode”.
Temp. / CO2
9.1 Idle mode
In this operating mode, temperature and CO
2
control are deactivated. The actual values continue to be
displayed in Normal Display. This operating mode is switched on or off via the controller function 1 “Idle
mode (chap. 9). In Normal Display the activated function 1 is indicated by an icon showing the number 1.
Normal display in Idle mode” (sample values)
Temp. / CO2