Binder CB-S 260 CO₂ incubator with hot air sterilization, series CB – 220-liter capacity

User Manual - Page 82

For CB-S 260.

PDF File Manual, 119 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB-S / CB-S-UL (E7) 06/2020 page 82/119
Replace the CO
2
sterile filter (once or twice a year depending on usage).
What to do in case of contamination?
Throw away / autoclave contaminated cultures.
Carefully inspect cultures that seem to be uncontaminated.
Clean the incubator as described. Wipe the inner chamber and the doors with a disinfectant and allow
to dry. Autoclave the shelves. Empty the water pan and autoclave it.
Perform hot air sterilization.
22. Cleaning, decontamination / disinfection, and sterilization
Clean the chamber after each use in order to prevent potential corrosion damage by ingredients of the
charging material.
Prior to renewed startup, allow the chamber to completely dry after all cleaning and decontamination
measures.
DANGER
Electrical hazard by water entering the chamber.
Deadly electric shock.
Do NOT spill water or cleaning agents over the inner and outer chamber surfaces.
Do NOT put ANY cleaning aids (cloth or brush) into slots or openings on the chamber.
Before cleaning, turn off the chamber at the main power switch (2) and disconnect the
power plug. Let the chamber cool down to ambient temperature.
Completely dry the chamber before turning it on again.
DANGER
Danger of explosion due to cleaning agents on hot surfaces.
Serious injury or death from burns and / or explosion pressure.
Before cleaning, turn off the chamber at the main power switch and disconnect the
power plug.
Before cleaning, let the chamber cool down to ambient temperature.
Completely dry the chamber before turning it on again.