Rigol DS8104-R oscilloscope

Rigol DS8000 R manual - Page 164

For DS8104-R.

PDF File Manual, 346 pages, Read Online | Download pdf file

DS8104-R photo
Loading ...
Loading ...
Loading ...
RIGOL Chapter 6 Operations and Measurements
6-36 DS8000-R User Guide
Click Measure → Statistic to enter the sub-menu of "Statistic".
âš« Click Statistic to enable or disable the statistical function.
âš« Click Reset Stat. to clear the history statistics data and makes statistics again.
âš« Click Count, then use the pop-up numeric keypad or scroll with the mouse to
set the count value. Its range is from 2 to 100000. By default, it is 1000.
All Measurement
All measurement could measure the time parameter, count value, voltage
parameters, and other parameters of the current measurement source and display
the results on the screen. Click Measure → All Measure to select "OFF" (disabling
the all measurement function) or select the desired channel for measurement
(CH1-CH4). If you select "CH1", data of all measurement parameters of CH1 are
displayed on the screen.
Note: The removal operation will not clear the measurement results of the
all-measurement function.
Cursor Measurement
Cursor measurement can measure the X-axis values (e.g. Time) and Y-axis values
(e.g. Voltage) of the selected waveform. Before making cursor measurement,
connect the signal to the oscilloscope to acquire stable display. All the parameters
supported by the "Auto Measurement" function can be measured with cursor
measurement. The cursor measurement function provides the following two cursors.
âš« X Cursor
X cursor is a vertical solid/dotted line that is used to make horizontal
adjustments. It can be used to measure time (s), frequency (Hz), phase (°), and
Y Cursor (Cursor B)
Y Cursor (Cursor A)
X Cursor (Cursor A)
X Cursor (Cursor B)
Loading ...
Loading ...
Loading ...