RIGOL Chapter 6 Operations and Measurements
6-28 DS8000-R User Guide
4
. Falling Edge Count: The number of falling edges that fall from above the
threshold upper limit to below the threshold lower limit.
Note:
âš« The above measurement items are only available for the analog channels.
âš« The default values for threshold upper limit and threshold low limit are 90% and
10%, respectively. You can click Measure in the function navigation menu, then
click Setting to enter the sub-menu to set the parameters.
Delay and Phase
Source A
Source B
Period
Delay
1. Delay A →B : indicates the time difference between the threshold middle
values of the rising edge of Source A and that of Source B. Negative delay
indicates that the rising edge of Source A occurred after that of Source B.
2
. Delay A →B : indicates the time difference between the threshold middle
values of the falling edge of Source A and that of Source B. Negative delay
indicates that the falling edge of Source A occurred after that of Source B.
Rising Edge Count = n
1 2 n
Falling Edge Count = n
1 2 n
Threshold Upper Limit
Threshold Lower Limit
Threshold Upper Limit
Threshold Lower Limit