Precision Loop Calibrator
Configuration Log and Data Log
33
The total number of data samples that can be logged:
709H - 9810 (less 2 x the number of sessions started and
stopped)
710 - 4910
Figure 48. Data Log Screen
The number of free records show on the first line. If data has
already been logged, the tag number shows below it.
The present battery voltage shows at the bottom to indicate if the
batteries should be changed before a log session starts. The log is
stopped before the Product is turned off when the battery reaches
its low voltage automatic shutoff limit of 5.6 V.
1. Use the selection knob to highlight and select the function.
2. Push the knob to do the operation.
3. Push
to return to the Log Select menu.
START operation:
If no free records or free sessions remain, or the present
HART device does not match the HART device already
logged, an error message shows.
Otherwise, proceed to interval selection described below.
ERASE operation:
If there is no logged data, an error message shows.
Otherwise, confirm that the present data is to be permanently
erased before doing so.
4. Select the logging interval. See Figure 49.
Figure 49. Logging Interval Screen
5. Turn the selection knob to move the highlight and select the
interval.
6. Push the selection knob to start logging at that interval.
7. Push
to go to the first data log screen.