Precision Loop Calibrator
Write LRV and URV Values
25
Write LRV and URV Values
Note
For the screens in this section, tag names can be
truncated to fit on one line. flashes in the upper-
right corner to show a live connection.
If the HART write commands are not active, these functions are
not available and an error message shows instead of the screen in
Figure 32.
Figure 32. Write LRV and URV Screen
Use the selection knob to choose the function. Push the selection
knob to choose the highlighted action and go to the corresponding
screen. See Write LRV and Write URV. Push
to go to the
Function Select menu.
Write LRV
The Product warns you to change the loop to MANUAL before you
proceed. Push the selection knob to proceed. Push
to go to
the Write LRV and Write URV menu.
The present LRV value and units show. See Figure 33.
Figure 33. Write LRV Screen
1. Push the selection knob to move the decade cursor.
2. Turn the selection knob to increment or decrement the value
in steps indicated by the selected decade.
3. Push and hold the knob to send the new value to the HART
device. An error shows if the HART device rejects the value.
4. Push
to go to the Function Select menu. A reminder to
change the loop to AUTOMATIC shows first.