RIGOL Chapter 13 Pass/Fail Test
13-2 DS8000-R User Guide
To Enable or Disable the Pass/Fail Test Function
Click on the function navigation icon
at the lower-left corner of the screen to
open the function navigation. Then, click on the "Pass/Fail" icon to open the
pass/fail test setting menu.
In the menu, click Enable continuously to enable or disable the pass/fail test
function.
Note: The pass/fail test function cannot be enabled in the following conditions:
when the time base mode is "XY" or "ROLL"; when the delay sweep is enabled.
To Start or Stop the Pass/Fail Test Operation
After the Pass/Fail test function is enabled, click Operate continuously to start or
stop the test operation.
During the test process, the oscilloscope will test the waveforms, display the test
information, and output the test failure information based on the current settings.
You can set the test mask, the display status of the test information, and the output
form of a failed test based on the selected source channel for the test. Then save the
test mask range to the internal or external memory, and then recall it when needed.
For the details, please refer to the following section.
Note:
âš« Only when the pass/fail test function is enabled, can you start or stop the
pass/fail test operation, enable or disable the display of the test information,
save and recall the test mask range.
âš« After starting the test operation, you can neither modify the source channel nor
adjust the test mask.
To Select the Source
Before selecting the source, connect the signal under test to the analog channel
input terminal of the oscilloscope. Click Source to select the desired source channel.
The available source channels include the analog channel (CH1-CH4).
Note: Only the channels that have been enabled currently can be selected.