SDS5000X Series Digital Oscilloscope User Manual
and may be less than memory depth. The actual sample points information can
be obtained in the timebase descriptor box (see the section "Timebase and
Trigger" for details).
The maximum memory depth in single channel mode is 2 times that of the dual
channel mode, as the following table:
Single Channel Mode
Dual Channel Mode
10k
5k
50k
25k
250k
125k
1.25M
625k
2.5M
1.25M
12.5M
6.25M
25M
12.5M
125M
62.5M
250M
125M
Single channel mode: Only one of C1/C2 is turned on, and only one of C3/C4
is turned on.
Dual channel mode: Both C1/C2 are turned on, or both C3/C4 are turned on.
Interpolation -- At small timebase settings, the number of original points on the
screen may be less than the number display pixels in the grid area, so
interpolation is necessary to display continuous waveform. For example, at 1
ns/div timebase and 5 GSa/s sample rate, the number of original points is 50,