SDS5000X Series Digital Oscilloscope User Manual
A. Set measurement threshold source
B. Set the type of threshold
C. Set the upper value
D. Set the middle value
E. Set the lower value
F. Return to previous menu
Threshold Type
Percent: Set according to the percentage of waveform. The setting range of
lower value and upper value is 1% ~ 99%, and the low value shall not be greater
than the middle value and high value
Absolute: Set according to the vertical range. The absolute threshold value
depends on the vertical scale, offset and probe attenuation. These values
should be set before setting absolute thresholds. Lower and upper values are
limited to the range of screen. If any absolute threshold is greater or less than
the minimum or maximum waveform value, the measurement may be invalid.