SDS5000X Series Digital Oscilloscope User Manual
dragging gesture or by entering the desired value in coordinate edit area.
For a side, the value is for its middle point. For a polygon, the ordinate is for its
geometric center.
Select a vertex (point)
Select a side (line segment)
Select a polygon
23.3 Pass/Fail Rule
The Pass/Fail rule is specified at the Type region in the Mask Test dialog box.
ï‚Ÿ All In: All data points must be inside the mask to pass the test. Even a
single point outside the mask will cause a failure.
ï‚Ÿ All Out: All data points must be outside the mask to pass the test. Even
a single point inside the mask will cause a failure.
ï‚Ÿ Any In: Any data point inside the mask will be recognized as pass. All
data points outside the mask will cause a failure.
ï‚Ÿ Any Out: Any data point outside the mask will be recognized as pass.