SDS5000X Series Digital Oscilloscope User Manual
condensation is expected.
IP Rating
IP20 (as defined in IEC 60529).
2.4 Cooling Requirements
This instrument relies on the forced air cooling with internal fans and
ventilation openings. Care must be taken to avoid restricting the airflow
around the apertures (fan holes) at each side of the scope. To ensure
adequate ventilation it is required to leave a 15 cm (6 inch) minimum gap
around the sides of the instrument.
CAUTION: Do not block the ventilation holes located on both sides of
the scope.
CAUTION: Do not allow any foreign matter to enter the scope through
the ventilation holes, etc.
2.5 Power and Grounding Requirements
The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC
power at 50/60 Hz (+/-5%).
No manual voltage selection is required because the instrument automatically
adapts to line voltage.