SDS5000X Series Digital Oscilloscope User Manual
16.5.2 CAN Trigger
Touch Trigger Setting in the
CAN trigger dialog box to
set the trigger condition:
 Start —The oscilloscope triggers at the beginning of the frame.
 Remote — The oscilloscope triggers on a remote frame with a specified ID.
ID, ID Bits (11-bit or 29-bit) and Curr ID Byte (1st, 2nd, 3rd, or 4th byte) can
be set. Curr ID Byte is used to specify the byte to be adjusted when using
the universal knob.
 ID —The oscilloscope triggers on the data frame that matches the specified
ID. ID, ID Bits (11-bit or 29-bit), and Curr ID Byte (1st, 2nd, 3rd, or 4th byte)
can be set.
 ID + Data —The oscilloscope triggers on the data frame that matches the
specified ID and data.ID, ID Bits (11-bit or 29-bit), Curr ID Byte (1st, 2nd,
3rd, or 4th byte), Data1 and Data2 can be set.
 Error—The oscilloscope triggers on the error frame.
16.5.3 CAN Serial Decode
The configuration of CAN decoding is similar to that of I2C decoding.