SDS5000X Series Digital Oscilloscope User Manual
value with the maximum
probability will be identified as
the base value; when set to min,
the minimum value of waveform
will be identified as the base
value
D. Return to previous menu
18.11 Threshold
Measurement thresholds can be defined by the user. This is more flexible than
fixed thresholds. For example, for pulse width measurement, the threshold can
be specified rather than fixed at 50%. For rise time, the lower / upper thresholds
can be specified rather than fixed at 10%/90%.
Changing the default threshold may change the measurement results of
relevant measurement items, such as: Period, frequency, +width, -width, +duty,
-duty, +BWidth, -BWidth, delay, T@M , rise time, fall time, CCJ, cycles, rising
edges, falling edges, edges, Ppulses, Npulses, and delay measurement.
Touch Config>Threshold in the measure dialog box to recall the threshold
dialog box: