SDS5000X Series Digital Oscilloscope User Manual
:AC coupling, 50Ω impedance
:Ground
Vertical Scale: The scale of each grid in the vertical direction. For example,
when the vertical scale is 1.00V/div, the full scale of the oscilloscope is
1.00V/div*8div=8V.
Vertical Offset: The offset of the channel in the vertical direction. When the
vertical offset is 0, the channel offset indicator is located in the middle of the
vertical axis.
Probe Attenuation Factor: Set the probe attenuation factor to match the
actual attenuation of the probe. The oscilloscope automatically calculates the
vertical scale according to the probe attenuation factor. For example, the
vertical scale of the oscilloscope is 100mV/div with 1X attenuation, and 1V/div
if the attenuation factor is changed to 10X. When inserting a standard 10X
passive probe with probe sense terminal, the oscilloscope will automatically
set the factor to 10X.
:1:1 attenuation, suitable for direct coaxial cable connection or passive
probes with 1X attenuation
:10:1 attenuation, suitable for general passive probes or active probes
with 10X attenuation
:100:1 attenuation, suitable for some high-voltage probes
:Custom attenuation factor