Loading ...
Loading ...
Loading ...
SDS5000X Series Digital Oscilloscope User Manual
Note: "Level +/-Delta” represents half of the actual window area. For example,
when the value is 200mV, it actually represents a range of ±200 mV, which is a
400mV window.
Holdoff, coupling and noise reject can be set in window trigger, see the sections
" Holdoff", " Trigger Coupling" and " Noise Reject" for details.
15.5.7 Interval Trigger
Trigger when the time difference between the neighboring rising or falling edges
meets the time limit condition.
When the trigger condition is set as an interval between two neighboring rising
edges and it is less than the set time value, the trigger diagram is as follows:
Trigger source, slope (rising, falling), limit range and time value can be set in
the trigger dialog box. Holdoff, coupling and noise reject can be set in interval
trigger, see the sections " Holdoff", " Trigger Coupling" and " Noise Reject" for
details.
Loading ...
Loading ...
Loading ...