SDS5000X Series Digital Oscilloscope User Manual
Setting Offset in the integral menu provides an approach to correct the DC
offset of the source. Small DC offsets in the input signal (or even small offset
errors of the oscilloscope itself) may cause the integral output waveform to
"ramp" up or down, as shown below:
Integral without Offset
Integral with Offset
In addition, the integral operation can be performed within a specified gate.
Touch Gate area in the math dialog box, and enable Gate function, then set
Gate A and Gate B to define the gate. The setting of the gate cursors is similar
to that of normal cursors. See " Select and Move Cursors" for details.